MENU
ES
  • EN
  • FR
  • ES
Search

7000 Calibrador de Temperatura RTD

7000 Calibrador de Temperatura RTD
  • 7000

Características

  • Exactitud 0.05°C (0.09°F) Resolución 0.01°C (0.02°F)
  • Exactitud 0.03Ω Resolución 0.01Ω
  • Medida a 2, 3, y 4 hilos
  • Mide y simula °C, °F, °K, y ohms
  • Rampa y pasos
  • PT100 mas otros 7 tipos de RTD
  • Programable por el usuario

7000 Imágenes

  • Un instrumento portátil de control de procesos que combina un termómetro digital de precisión (usando sondas RTD) con un calibrador de RTD/ohms.

  • Modes of Operation - Detailed Specifications

    Monitor Mode (4 wire)

    Typical Functions

    1. Check RTD probes by measuring their resistance at known temperatures.
    2. Measure resistance values.

    3. Indicate temperature when connected to an RTD probe.

    4. Can be pre-programmed with a particular RTD’s characteristics to allow very high accuracy.

    Excitation current 1mA on all ranges
    Resistance range 0.01Ω to 2.6kΩ
    Resolution 0.01Ω
    Accuracy See specifications below
    Auto re-calibration Every 0.6 secs
    Temperature stability Better than 0.0015% per °C
    Max/Min values Logged automatically

    The 7000 may be used with a calibrated and certified probe to produce a highly accurate thermometer.
    The performance can be further enhanced by programming the actual characteristic of the probe into the unit.

    Simulator Mode (4 wire)

    Typical Functions

    1. Output resistance of precise known value.

    2. Simulate an RTD value from an RTD table chart.

    3. Simulate an RTD value using the internal table.

    Excitation current 0.6mA to 1mA
    Resistance range 0.01Ω to 2.6kΩ
    Resolution 0.01Ω
    Accuracy See specifications below
    Auto re-calibration Every 0.6 secs.
    Temperature stability Better than 0.0015% per °C

    Enhanced performance may be achieved by programming the unit to simulate the characteristic of a particular probe.
    Five fixed step points (0, 25, 50, 75, 100%) are available between a user set minimum (0%) and a maximum (100%).
    Programmable ramp function is also available.


    Technical Specifications

    Standard RTD types (Non standard RTD types user programmable)
    Element Alpha Coefficient Temp Range °C Accuracy °C Temp Range °F Accuracy °F

    Pt100 DIN

    0.003850

    -200 to 250
    250 to 849

    0.05
    0.07

    -330 to 480
    480 to 1560

    0.10
    0.14

    Pt100 US

    0.003916

    -100 to 250

    250 to 457

    0.05

    0.07

    -150 to 480

    480 to 850

    0.10

    0.14

    Pt200 DIN 0.003850 -200 to 300 0.05 -330 to 570 0.10

    Pt500 DIN

    0.003850

    -200 to 250

    250 to 630

    0.05

    0.07

    -330 to 480

    480 to 1160

    0.10

    0.14

    Pt1000 DIN

    0.003850

    -200 to 250

    250 to 630

    0.05

    0.07

    -330 to 480

    480 to 1160

    0.10

    0.14

    Ni 120 0.006180 -100 to 200 0.05 -150 to 390 0.10
    Ni 1000 0.006180 -100 to 200 0.05 -150 to 390 0.10
    Resistance Accuracy
    Range (Ohms) Monitor Generator
    20 to 400 0.03Ω 0.03Ω
    400 to 800 0.10Ω 0.10Ω
    800 to 1200 0.20Ω 0.20Ω
    1200 to 2600 0.50Ω 0.50Ω*

    * plus additional error of 0.05% of output value. If excitation current is less than 1mA.


    General Specifications

    Operating temp range -10 to 50°C

    Battery power

    NiMH rechargeable

    Mains power

    External mains adaptor (optional)

    Battery life > 24hrs typical use
    Case Impact resistant ABS
    Dimensions H165 x W90 x D45mm
    Weight 0.42kg
    Supplied With Carry case, mains charger, pair of test leads, and Manual
    Optional Extras Mains adaptors (230 or 110V AC) and calibration certificates traceable to NPL and UKAS
    Country of Origin UK
  • Modelos

    7000

    Calibrador de Temperatura RTD

    Opciones
    7633 Mains Adaptor (230 or 110V AC must be specified on order)
    C183 Certificado de calibración fábrica
    C194 Certificado de calibración UKAS (ISO 17025)
  • Descargas PDF
    Nombre del archivo Tamaño del archivo Idioma

    7000 Data sheet

    277kB

    EN

  • Solicite una cotización

    7000 Calibrador de Temperatura RTD

    Opciones del producto
    1.20

Productos relacionados

  • 1049 Simulador PT100

    • -200°C a + 800°C
    • 23 puntos de ajuste
    • Exactitud de  ± 0.3°C
    • Coeficiente de temp. < 30ppm/°C
    • ITS-90 IEC60751
    • Excede clase A
    • Fuente de Resistencia pasiva
    Ver 1049
  • 1067 Caja de Decadas de Resistencia de Precision

    • 10mΩ to 12kΩ
    • Exactitud de 0.01% 
    • Resistencia residual: < 10mΩ
    • Potencia Máx: 0.35 watt por resistor
    • 20ppm/año (>1Ω)
    • Máximo 200V DC/AC RMS
    • Coeficiente de temp:10ppm/ºC
    Ver 1067
  • 1090 Calibrador de Procesos y Temperatura

    • Mide/Simula termopares
    • Mide/Simula PT100, mV y mA
    • Unidades en °C, °F, μV/mV, mA
    • Exactitud: Fuente 0.02%/0.5°C
    • Exactitud: Medición 0.05%/0.7°C
    • Automático CJC - seleccionable
    • Funcionamiento a batería o red
    Ver 1090
  • 5011 Calibrador de Resistencia/Temperatura

    • 1Ω - 120MΩ
    • Simulación RTD
    • Simulación de termocoples
    • 0 a 22V DC tensión - opción
    • Opción de corriente DC
    • Opción de frecuencia de 10MHz
    • Control por PC vía software EasyCal
    Ver 5011
  • 7005 Calibrador de Tensión/Corriente/Lazo

    • Medicion: 0 - 25V or 0 - 125mA
    • Fuente: 0 - 21V or 0 - 50mA
    • Exactitud: 0.01%
    • Resolución: 1μA o 1mV
    • Funciones de transmisor y raíz cuadrada.
    • Pasos y rampa programable
    • Auto-rango
    Ver 7005