Caractéristiques Principales
- Précision 0,09°F, Résolution 0,02°F
- Précision 0,03Ω, Résolution 0,01Ω
- Connexions 2, 3 et 4 fils
- Mesure et simule °C, °F, °K et ohms
- Echelons et rampe
- PT100 + 7 autres types RTD
- Utilisateur programmable
Instrument de contrôle de process portable combinant un thermomètre numérique de précision (sondes RTD) avec un calibrateur de résistance / RTD.
Modes of Operation - Detailed Specifications
Monitor Mode (4 wire) Typical Functions
1. Check RTD probes by measuring their resistance at known temperatures.
2. Measure resistance values.3. Indicate temperature when connected to an RTD probe.
4. Can be pre-programmed with a particular RTD’s characteristics to allow very high accuracy.
Excitation current 1mA on all ranges Resistance range 0.01Ω to 2.6kΩ Resolution 0.01Ω Accuracy See specifications below Auto re-calibration Every 0.6 secs Temperature stability Better than 0.0015% per °C Max/Min values Logged automatically The 7000 may be used with a calibrated and certified probe to produce a highly accurate thermometer.
The performance can be further enhanced by programming the actual characteristic of the probe into the unit.Simulator Mode (4 wire) Typical Functions
1. Output resistance of precise known value.
2. Simulate an RTD value from an RTD table chart.
3. Simulate an RTD value using the internal table.
Excitation current 0.6mA to 1mA Resistance range 0.01Ω to 2.6kΩ Resolution 0.01Ω Accuracy See specifications below Auto re-calibration Every 0.6 secs. Temperature stability Better than 0.0015% per °C Enhanced performance may be achieved by programming the unit to simulate the characteristic of a particular probe.
Five fixed step points (0, 25, 50, 75, 100%) are available between a user set minimum (0%) and a maximum (100%).
Programmable ramp function is also available.
Technical SpecificationsStandard RTD types (Non standard RTD types user programmable) Element Alpha Coefficient Temp Range °C Accuracy °C Temp Range °F Accuracy °F Pt100 DIN
0.003850
-200 to 250
250 to 8490.05
0.07-330 to 480
480 to 15600.10
0.14Pt100 US
0.003916
-100 to 250
250 to 457
0.05
0.07
-150 to 480
480 to 850
0.10
0.14
Pt200 DIN 0.003850 -200 to 300 0.05 -330 to 570 0.10 Pt500 DIN
0.003850
-200 to 250
250 to 630
0.05
0.07
-330 to 480
480 to 1160
0.10
0.14
Pt1000 DIN
0.003850
-200 to 250
250 to 630
0.05
0.07
-330 to 480
480 to 1160
0.10
0.14
Ni 120 0.006180 -100 to 200 0.05 -150 to 390 0.10 Ni 1000 0.006180 -100 to 200 0.05 -150 to 390 0.10 Resistance Accuracy Range (Ohms) Monitor Generator 20 to 400 0.03Ω 0.03Ω 400 to 800 0.10Ω 0.10Ω 800 to 1200 0.20Ω 0.20Ω 1200 to 2600 0.50Ω 0.50Ω* * plus additional error of 0.05% of output value. If excitation current is less than 1mA.
General SpecificationsOperating temp range -10 to 50°C Battery power
NiMH rechargeable
Mains power
External mains adaptor (optional)
Battery life > 24hrs typical use Case Impact resistant ABS Dimensions H165 x W90 x D45mm Weight 0.42kg Supplied With Mallette de Transport, mains charger, pair of test leads, and Manuel Optional Extras Mains adaptors (230 or 110V AC) and calibration certificates traceable to NPL and UKAS Country of Origin UK Modèles 7000
Calibrateur de Température RTD
Options 7633 Mains Adaptor (230 or 110V AC must be specified on order) C183 Certificat de Calibration Usine C194 Certificat de Calibration UKAS (ISO 17025) Téléchargements PDF Nom du fichier Taille du fichier Langue 277kB
EN
Appel d'offre
7000 : Calibrateur de Température RTD
Produits connexes
-
1049 : Simulateur PT100
- -200°C à +800°C
- Précision : ± 0,3°C
- ITS-90 IEC60751
- 23 points de réglage
- Supérieur à la classe A
- Bonne stabilité
- Source de résistance passive
-
1067 : Boîte à décades de résistances de précision
- 10mΩ à 12kΩ
- 6 décades
- Précision: 0,01%
- Résistance:Inférieure à 10mΩ
- Puissance: 0,35W par résistance
- Bonne stabilité à long terme
- Tension: 200V DC/AC RMS
-
1090 : Calibrateur de Process
- Mesure / Simule thermocouples
- PT100, RTD, mV et mA
- °C, °F, µV/mV, mA
- Boucle de tension 24V
- 4 à 20mA et 0 à 50mA
- Alimentation batterie ou secteur
- Mallette de Transport
-
5011 : Calibrateur de Résistance / Température
- 1Ω à 120MΩ
- Simulation RTD
- Simulation thermocouple en option
- Fréquence jusqu’à 10MHz en option
- Interfaces RS232/GPIB/USB
- Fonctions accessibles en face avant
- Contrôle via logiciel EasyCal
-
7005 : Calibrateur tension / Courant / Boucle
- Courant: 125mA
- Tension: 25V
- Simulation: 50mA ou 21V
- Précision : 0,01%
- Résolution 1μA ou 1mV
- Rampe et échelons programmables
- Changement gamme automatique