1050 PT100 Simulator

1050 PT100 Simulator

Features

  • -100°F to + 1000°F
  • ± 0.5°F accuracy
  • Based on ITS 90 EN60751
  • 23 set points
  • Exceeds class A
  • Good temperature stability
  • Passive resistance source
 

The 1050 is a handheld precision simulator for PT100 0.3850 platinum resistance elements used for accurate temperature measurement.

It follows the PT100 scale from -100°F to +1000°F with 23 set points. High performance metal film resistors are used throughout which ensures a good temperature coefficient and long term stability.

The specification is in accordance with DIN EN 60751 (ITS 90). Offering high accuracy across the full operating range of PT100 devices it exceeds the performance of Class A & B. The 1050 will be of particular interest to those operating in the -80 to +140°F range where a performance exceeding Class A (e.g. better than +/- 0.3°C at 0°F), is required.

Since the 1050’s output is a purely passive resistance it will operate with all types of PT100 measuring equipment including the live systems using pulsed, or interrupted excitation current. The pocket sized design (112 x 61 x 55mm) makes it easily portable and ideal for lab or field use. The instrument is supplied as standard with a carry case.

This 1050 is a popular test instrument in the Time Electronics process calibration equipment range. As a PT100 simulator for temperature measurement equipment it offers high stability and ease of use across applications..

Technical Specifications

Set Points °F -100, -50, -25, 0, 32.2, 50, 75, 100, 125, 150, 175, 200, 225, 250, 300, 350, 400, 450, 500, 600, 700, 800, 1000
Range -100 to 300°F 300 to 1000°F
Accuracy ± 0.5°F ± 0.9°F
Temperature Coefficient Less than 30ppm/°F
Maximum Current 50mA


General Specifications

Dimensions H112 x W61 x D55mm (2.4 x 5 x 2.2”)
Weight 0.17kg (0.4lb)
Supplied With Carry case and user manual
Optional Extras Calibration certificates traceable NPL and UKAS
Country of Origin UK
Models

1050

PT100 Simulator (°F)

1049

PT100 Simulator (°C) - View web page

Options

C161

Factory Calibration Certificate (NPL traceable)

C114 UKAS Calibration Certificate (ISO 17025)
PDF Downloads
File Name File Size Language

1050 Data Sheet

170kB

EN

1049 (°C version) User Manual

277kB

EN / FR / DE / IT