5069 Insulation Tester Calibrator

5069 Insulation Tester Calibrator

Features

  • 100KΩ to 100GΩ
  • Basic accuracy 1%
  • Up to 10kV operation
  • Battery operation
  • Fully shrouded safety connectors
  • Display of open circuit voltage
  • Display of short circuit current
 

A precision instrument suitable for calibrating and testing general purpose insulation testers and megohmmeters with test voltages up to 10kV.

It is constructed in a high strength co-polymer plastic case and is powered by a rechargeable battery. This ensures full isolation from the mains and prevents stray leakage.

The insulation tester being calibrated can be tested for open circuit voltage and short circuit current. These are displayed on the digital meter mounted on the front panel. The insulation resistance is provided by a precision 4 dial decade resistance bank which can be set to a maximum of 99.99GΩ with additional resistance values of 100kΩ, 200kΩ, 500kΩ, 1MΩ, 2MΩ and 5MΩ which can be switched in as required.

Rugged and portable, the 5069 is ideal for site calibration work. Also available is the 5068 InsCal, 2.5kV version.

Technical Specifications

Resistance 10GΩ, 1GΩ, 100MΩ, 10MΩ decade ranges.
Fixed Values 5MΩ, 2MΩ, 1MΩ, 500KΩ, 200KΩ, 100KΩ.
Resistance Accuracy 100kΩ to 5MΩ = 1%. 10MΩ to 10GΩ = 1%. 10GΩ to 100GΩ = 5%.
Resistance Temperature Coefficient Less than 250ppm per °C
Open Circuit Voltage Measure 0 to 2kV range, accuracy 1% of f/s. 0 to 10kV range, accuracy 2% of f/s.
Voltage Display 1.999kV full scale, and 10kV full scale.
Short Circuit Current Measure 0 to 2mA range and 0 to 20mA range = 1% of f/s accuracy.
Current Display 1.999mA full scale, and 19.99mA full scale.


General Specifications

Power Internal battery, 6V re-chargeable NiCad, >150hrs between charges.
Dimensions W406 x H175 x D330mm
Weight 4.4kg
Supplied With Safety connection leads with bare end to allow custom probes (10kV rated), 4mm 5kV test lead set, wallet to carry test leads, and mains charger.
Optional Extras Calibration certificates traceable NPL and UKAS
Country of Origin UK
Models

5069

INSCAL: Insulation Tester Calibration System (100kΩ to 100GΩ, up to 10kV operation)

5068

INSCAL: Insulation Tester Calibration System (100kΩ to 10GΩ, up to 2.5kV operation). View web page

Options

C189

Factory Calibration Certificate (NPL traceable)

C112 UKAS Calibration Certificate (ISO 17025)
PDF Downloads
File Name File Size Language

5069 Data Sheet

215kB

EN

5069 User Manual

325kB

EN

5068 Data Sheet

215kB

EN