5065 6.5 digit Multimeter

5065 6.5 digit Multimeter

Features

  • Resolution 6 1/2 digits
  • Accuracy: 0.005% DC voltage
  • USB interface
  • Optional GPIB interface
  • Temperature measurements
  • 10 channel scanner option
  • EasyCal software compatible
 

The 5065 is a high performance, versatile 6½ digit bench multimeter with 19 measurement functions. 

Low cost, simple operation, stability, and high accuracy make the 5065 an ideal DMM for a variety of applications. In addition high speed means both the sampling rate and the data transfer rate can achieve 2000 readings per second.

The comprehensive range of features makes the 5065 suitable for test engineers, R&D, service, and calibration technicians. The durable construction with a protective rubber guard and carry handle allows convenient bench top use, and transportability.

The front panel features a two-line, colour-coded display for easy to read measurements and functions. Control is simple, with common functions selected from a single button press.

The 5065 measures AC/DC voltage, AC/DC current, 2/4 wire resistance, frequency, period, diode, continuity, thermocouples and RTD.

In addition to standard functions many other capabilities such as Min/Max, Ratio/% and Null are included. Memory functions with up to 2000 readings can be stored and recalled. Limit testing with an external output can be configured to signal pass or fail.

The USB interface enables connectivity to a PC. A basic setup and graphing application is supplied as standard. Add-in utilities for Microsoft Excel and Word allow simple data collection directly into a spreadsheet or document.

An optional 10 channel scanner card is available which allows multiple inputs to be measured. Further options are test lead sets, GPIB and RS232 interfaces, and an RTD probe adaptor.

Calibration made easy
Connect the 5065 to a PC/Laptop installed with Time Electronics’ EasyCal software and automate the calibration process. This provides increased speed of calibration and consistency of results. Easily produce calibration certificates and reports to ISO 9001, ISO 17025, and other international quality standards.

Technical Specifications

DC Voltage
Range Resolution Input Resistance Accuracy
100.0000mV 0.1μV > 10GΩ 0.0050 + 0.0035
1.000000V 1.0μV > 10GΩ 0.0040 + 0.0007
10.00000V 10μV > 10GΩ 0.0035 + 0.0005
100.0000V 100μV 10MΩ 0.0045 + 0.0006
1000.000V 1μV 10MΩ 0.0045 + 0.0010
DC Current
Range Resolution Shunt Resistance Accuracy
10.000000mA 10nA 5.1Ω 0.05 + 0.02
100.00000mA 100nA 5.1Ω 0.05 + 0.005
1.000000A 1μA 0.1Ω 0.1 + 0.01
3.00000A 10μA 0.1Ω 0.12 + 0.02
AC Voltage
Range Resolution Frequency (Hz) Accuracy

100.0000mV

0.1μV

3 to 5
5 to 10
10 to 20K
20 to 50K
50K to 100K
100K to 300K

1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50

1.000000V
to
750.000V

1.0μV to 1mV

3 to 5
5 to 10
10 to 20K
20 to 50K
50K to 100K
100K to 300K

1.00 + 0.03
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.60 + 0.08
4.00 + 0.50

AC Current
Range Resolution Frequency (Hz) Accuracy

1.000000A

1μA

3 to 5
5 to 10
10 to 5K

1.00 + 0.04
0.30 + 0.04
0.10 + 0.04

3.000000

10μA

3 to 5
5 to 10
10 to 5K

1.10 + 0.06
0.35 + 0.06
0.15 + 0.06

Resistance
Range Resolution Test Current Accuracy
100.0000Ω 100μΩ 1mA 0.01 + 0.004
1.000000kΩ 1mΩ 1mA 0.01 + 0.001
10.00000kΩ 10mΩ 100μA 0.01 + 0.001
100.0000kΩ 100mΩ 10μA 0.01 + 0.001
1.000000MΩ 5μA 0.01 + 0.001
10.00000MΩ 10Ω 500nA 0.04 + 0.001
100.0000MΩ 100Ω 500nA 0.8 + 0.01

Diode Test
1.0000V


10μV


1mA


0.01 + 0.02

Continuity
1000.00KΩ


10mΩ


1mA


0.01 + 0.03

Frequency and Period
Range Frequency (Hz) Accuracy (% of reading)

100mV to 750V

3 to 5
5 to 10
10 to 40
40 to 300K

0.1
0.05
0.03
0.06

Temperature
RTD Types PT100, D100, F100, PT385 or PT3916. (Best Accuracy ±0.08°C )
Thermocouple Types J, K & T (Accuracy ±0.5°C ). E (Accuracy ±0.6°C ). R, S & B (Accuracy ±5°C )


General Specifications

Standard Interfaces RS-232 and USB (GPIB optional)
Line Power 100 - 230V AC 50/60Hz
Dimensions W210 x H85 x D350mm
Weight 4.4kg

For a full list of options please see the Models/Options tab.

Models

5065

Bench Digital Multimeter

Options
9714 10 Channel Scanner card
9715 GPIB Interface (replaces standard fitted RS232 interface)
9717 RTD Probe Adaptor
9713 Carry Case with test lead compartment

9541

Basic Test Lead Set - View data sheet

9796

Premium Test Lead Set - View data sheet

C173 Factory Calibration Certificate (NPL traceable)
C117 UKAS Calibration Certificate (ISO 17025)
Calibration Software and Accessories

ECFLA

EasyCal Full License - when purchased with multifunction calibrator/DMM: Discounted full license with 1 year support. 
For further licenses and EasyCal add-ons see the Easycal web page or PDF brochure

9795 Printer & Connectivity Kit (Printer, DVD-RW, 4 Port USB Hub, Numeric Key Pad & USB Memory Stick)
9777 Bar Code Reader
9778 Cal and ID Label Printer

9779

Job and Address Label Printer

ESP1 EasyCal Support Package 1: 1 year email and telephone support. Minor upgrades.
ESP2 EasyCal Support Package 2: 2 year email and telephone support. Minor upgrades.
EOT1 EasyCal Online Training Session (Via Remote Desktop)
PDF Downloads
File Name File Size Language

5065 Data Sheet

370kB

EN

EasyCal Brochure

3MB (Low resolution)

EN

EasyCal Brochure

7MB (Higher resolution)

EN

Test Lead Sets Data Sheet

172kB

EN