The 1007 is a handheld calibrator that can be used for potentiometric voltage measurement in addition to its operation as a millivolt source.
The null zero and sensitivity are adjustable via front panel controls. Maximum sensitivity enables null balance to resolve 3 microvolt.
Three output ranges are provided to give adjustable output values from 1μV to 1V with a basic 0.02% accuracy. For signal injection, the operator needs to switch on, check the battery condition, select the range, and set the required voltage using the thumbwheel switches. The 1007 uses a precision reference diode and low temperature coefficient resistors to give a highly stable output.
Power is provided by 6 AA batteries. Battery life is several months, depending on usage. The battery condition is monitored by an indicator mounted on the top of the unit. The 1007 has up to 20mA drive current and is short circuit and overload protected. An off/normal/reverse output polarity switch is provided.
Safety Terminals
Fitted as standard and fully compatible with 4mm shrouded plugs, as well as standard plugs, bare wires, and spade terminals.
Added Protection
The 1007 comes fitted with an ergonomic rubber cover providing increased protection and durability. It has a textured grip for comfortable handling and openings at the top and bottom to allow access to the battery meter and a position to place labels if required. It is easy to remove if the user prefers a stand-alone unit or to house the 1007 in the 9027 carry case.
Applications
Suitable for calibration and simulation of thermocouples. Accurate voltages equivalent to the output from a thermocouple can easily be set on the 1007, enabling fast calibration of temperature measuring equipment.
Alternatively, the 1007 can measure thermocouples output by operating as a potentiometer. Other applications include chart recorder calibration, A/D converter and DMM calibration, and use as a stable voltage for backing off DC offsets.
The 1007 is an easy to use test instrument designed for accuracy and reliability. As part of the Time Electronics portable calibration equipment range it is ideal for use in the lab or field.
Output |
0 to 999.9mV in 3 ranges: |
Accuracy |
± 0.02% of setting + ± 0.02% of range + ± 1µV. |
Output Resistance |
Less than 0.2Ω on 1 V and 100 mV ranges. 1Ω on 10mV range. |
Maximum Output Current |
1V and 100mV ranges = 20mA. 10mV range = Up to short circuit value although it should be noted that loads of less than 1kΩ will give greater than 0.1% error. |
Output Voltage Stability |
Less than 60ppm/°C. Less than 100ppm per 3 months (non-cumulative). |
Output Polarity |
Positive or negative switch selected. A centre ‘off’ position is also provided. |
Output Noise Level |
Less than 30ppm of full scale |
Reference Source |
Precision zener diode, selected for stability and low temperature coefficient. |
Maximum Overload |
The instrument can withstand continuous short circuit on the output for all ranges. |
Null Balance Display |
On a front panel meter, zero and sensitivity controls are provided: |
Power Supply |
6-AA size (51x14mm) batteries. A battery condition display indicates when the batteries should be changed. An alternative power source is 6 NiMH cells of the same dimensions. These can be recharged via a socket on the top of the unit. The 6 rechargeable batteries and mains recharger are available as an optional extra. |
Operating Temperature |
-10°C to +60°C |
Dimensions |
H200 x W75 x D110mm (H215 x W100 x D120mm including protective cover) |
Weight |
0.75kg (1.2kg including protective cover) |
| Supplied With | Protective cover (fitted), 6 x AA batteries (installed), and user manual |
Optional Extras |
Carry case, rechargeable battery pack (240V or 110V mains), and calibration certificates traceable to NPL and UKAS |
| Models |
|---|
1007 |
DC Millivolt Potentiometer and Source |
| Options |
|---|
9027 |
Leatherette Carry Case |
9529 |
Rechargeable Battery Pack - (6 NiMH Cells + 230V Mains Charger) |
| 9528 | Rechargeable Battery Pack (8 NiMH Cells + 110V Mains Charger) |
C150 |
Factory Calibration Certificate (NPL traceable) |
C101 |
UKAS Calibration Certificate (ISO 17025) |
Also available: 1006 DC Millivolt Source