7050 Multifunction Process Calibrator

7050 Multifunction Process Calibrator
  • 7050 Multifunction process calibrator

Features

  • Measure to 40V, 400mA, 40kΩ
  • Output voltage up to 10V
  • Output current to 20mA
  • Output resistance to 40kΩ
  • Accuracy 0.03% of range
  • Thermocouple measure & simulate
  • RTD measure & simulate: PT100

7050 Images

  • The 7050 multifunction process calibrator combines the essential functions of measurement and simulation of volts, millivolts, milliamps, and ohms with the direct readout simulation of thermocouples and RTD’s in °C or °F units.The instrument is microprocessor based and housed in a rugged case designed for use in the field. Input and Output functions are displayed simultaneously and the switch controls allow for easy and precise operation.

    Five calibration points are instantaneously accessible through a preset percentage selector and all functions may be generated in a fully programmable ramp format.In addition to the standard calibration facilities the instrument may be used as a signal converter. Any electronic input signal can be converted and output as a proportional signal of 4-20mA, 0-5V or sq. root 4-20mA.

    The 7050 is highly adaptable, IPT 68 and ITS 90 curves are accommodated and non-standard thermocouples and RTD’s may be programmed on request.

  • Technical Specifications

    Voltage / Current / Resistance Generation and Measure 

    INPUT

    Voltage
    Range Resolution Accuracy
    40V 1mV 0.03%
    4V 100μV 0.03%
    400mV 10µV 0.03%
    40mV 1µV 0.03%
    Current
    Range Resolution Accuracy
    400mA 10μA 0.03%
    40mA 1μA 0.03%
    4-20mA/0-100% 0.01% 0.06%
    Resistance
    Range Resolution Accuracy
    40000Ω 0.03%
    4000Ω 0.1Ω 0.03%
    400Ω 0.01Ω 0.06%

     

    OUTPUT

    Voltage
    Range Resolution Accuracy
    -2 to +10V 1mV 0.06%
    -100mV to +400mV 10μV 0.03%
    -10mV to +40mV 1µV 0.03%

    Output Impedance: < 10Ω

    Current
    Range Resolution Accuracy
    0 to 20mA 10μA 0.03%
    4 to 20mA 1μA 0.03%
    TX Sim (4 to 20mA) 0.01% 0.03%

    Max Load Resistance: 900Ω at 20mA, Open circuit voltage 18- 27V 

    Resistance
    Range Resolution Accuracy
    0 to 400Ω 0.01Ω ± 0.03%

    Excitation Current 0.2 to 2mA 

    Other Specifications: Loop drive supply: Nominal 24V, current limit set to 30mA


    Process Signal and Temperature Measurement & Simulation

    Temperature
    Thermocouple Type Temp Range °C Accuracy °C

    J

    -210 to 0
    0 to 710
    710 to 1200

    ± 0.4 
    ± 0.2 
    ± 0.9

    K

    -270 to -240 
    -240 to -180 
    -180 to -50 
    -50 to 960 
    960 to 1370

    ± 3.0 
    ± 1.2 
    ± 0.5 
    ± 0.3 
    ± 1.7

    T

    -270 to -250 
    -250 to -180 
    -180 to -100 
    -100 to 400

    ± 3.0 
    ± 1.0 
    ± 0.5 
    ± 0.2

    R

    -50 to 0 
    0 to 70 
    70 to 400 
    400 to 1000 
    1000 to 1760

    ± 3.0 
    ± 2.0 
    ± 1.5 
    ± 1.0 
    ± 0.8

    S

    -50 to 0 
    0 to 100 
    100 to 450 
    450 to 1760

    ± 2.5 
    ± 1.8 
    ± 1.3 
    ± 0.9

    B

    200 to 400 
    400 to 1000 
    1000 to 1820

    ± 5.0 
    ± 2.0 
    ± 1.0

    N

    0 to 330 
    330 to 1100 
    1100 to 1300

    ± 0.4 
    ± 0.3 
    ± 1.5

    E

    -230 to 0 
    0 to 540 
    540 to 1000

    ± 0.5 
    ± 0.2 
    ± 1.5

    RTD - PT100
    Range °C Accuracy °C
    -200 to 850 ± 0.3
    Thermocouple Linearisation to BS 4937. ITS 90 or IPTS68 selectable.
    Pt100 Linearisation to BS1904 (1984).DIN43760 (1980). Linearisation Acc: 0.1% of reading for Class A, 38.5Ω F.I
    Excitation current 0.5mA (Input), 200μA to 2mA (Output).
    CJC External (Pt100), Internal, Ice point. Accuracy +/-0.2°C at 23°C
    Ramp Fully programmable continuous and stepped
    Steps 5 Steps with fully adjustable zero
    Transmitter Tx function 4-20mA
    Signal Converter Any input to any output
    Dual Readout Measure and source simultaneously displayed


    General Specifications

    Stability 100ppm per °C range 0 to 40°C
    Displays LCD Dot Matrix, 1 line by 20 charx12mm, alphanumeric.

    Dimensions / Weight

    W270 x H175 x D250mm, weight 4.5kg

    Optional Extras Calibration Certificates traceable to NPL and UKAS
  • Models

    7050

    Process and Thermocouple Calibrator

    Options
    C180 Factory Calibration Certificate (NPL traceable)
    C192 UKAS Calibration Certificate (ISO 17025)
  • PDF Downloads
    File Name File Size Language

    7050 Data Sheet

    312kB

    EN

  • Request for Quotation

    7050 Multifunction Process Calibrator

    Product Options
    1.20

Related Products

  • 1090 Portable Process Calibrator

    • Measure/simulate thermocouples
    • Measure/simulate PT100
    • Measure/source (µV/mV/mA)
    • Display in °C and °F
    • Automatic CJC - selectable
    • 10 point memory recall
    • Process loops 4-20mA/0-50mA
    View 1090
  • 7000 RTD Temperature Calibrator

    • 0.05°C accuracy, 0.01°C resolution
    • 0.03Ω accuracy, 0.01Ω resolution
    • 2, 3 and 4 wire connections
    • Read and simulate °C, °F, °K, Ω
    • Ramp and step
    • PT100 plus 7 other RTD types
    • User programmable
    View 7000
  • 7005 Process Loop Calibrator

    • Loop source/measure - V or I
    • Measure 0 - 25V or 0 - 125mA
    • Source 0 - 21V or 0 - 50mA
    • Accuracy 0.01%
    • Resolution 1μA or 1mV
    • Programmable ranges
    • Manual step, auto-step & ramp
    View 7005
  • 7006 Loop Simulator & Source

    • 4 - 20mA or 0 - 10V loops
    • 7 calibration set-points
    • Accuracy 0.1%
    • TxSim and RxTest modes
    • Autostep output - adjustable rate
    • Battery powered PP3
    • Internal loop drive supply
    View 7006
  • 7007 Loop Signal Indicator

    • 4 - 20mA, 0 -10V, 0 - 50V ranges
    • LCD 4 digit display, mA, V, % of rng
    • Accuracy 0.05%
    • RxSim, TxTest, or 50mA/50V modes
    • Internal loop supply, 25mA max
    • Battery powered 9V PP3
    • Supplied with carry case 
    View 7007