7077 Process Calibrator Module

  • The 7077 is a high performance process calibrator module that combines source and measurement functions for thermocouples, PT100s, μV, mV, and mA. As a multifunction module the 7077 combines accuracy with simple operation, making it ideal for process plant applications.

    Thermocouple Measurement and Simulation
    Measure and simulate the temperature and mV characteristics of J, K, T, R, S, B, N and E thermocouples.

    Cold Junction Compensation
    The unit can be operated with or without internal cold junction compensation.

    PT100 Measurement and Simulation
    Based on 0.3850 alpha probe standard. Range is -200°C to 700°C.

    Measurement and Source (μV, mV, and mA)
    Measurement ranges are 0 to +/-30mV and 0 to +/-60mA. Source ranges are 0 to +/-80mV and 0 to +80mA.

    Temperature Units Selection
    The display can be easily changed from °C to °F. The equivalent μV (thermocouples) and ohms (PT100) can also be shown.

    24V Process Loop Drive Mode
    A process loop can be driven at 24V and up to 60mA by selecting the 'Milliamp Source' mode and setting it at 60mA (or a lower level if required).

    Inching (Incrementing/Decrementing)
    The 7077 has a general-purpose inching function. This adjusts the output in fixed increments of temperature (thermocouples only) or voltage or current. The set-up menu gives a the user a choice of three levels of increment i.e. 0.1, 1 or 10 for °C/°F, or 1, 10, or 100 μV/μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. This is especially useful for calibrating thermostat controllers that have tight specification on hysteresis.

    Memory Recall and Step/Auto-Step Functions
    Up to 10 values can be stored in the unit's non- volatile memory and they can be recalled at any time. The user can also manually step through them in sequence using the step key. Continuous stepping (auto-step) is also available at any user selectable rate between 1 and 10 seconds/step. 

  • temperature measure
    Thermocouple Type Temperature Range (°C) Accuracy (°C)
    J -200 to 580 0.7

    K

    -200 to -150
    -150 to 750

    2.5
    0.5

    T

    -200 to 0
    0 to 400

    1.5
    0.4

    R

    50 to 400
    400 to 1750

    3.0
    1.5

    S

    -50 to 100
    100 to 1750

    3.0
    1.5

    B

    110 to 1000
    1000 to 1800

    3.5
    1.5

    N -100 to 890 0.6
    E -50 to 400 0.4
         

    Resolution

    0.1°C or 0.1°F

     
    Temperature Simulation
    Thermocouple Type Temp Range °C Accuracy °C

    J

    -210 to 150
    150 to 1200

    0.15
    0.3

    K

    -270 to 190
    190 to 1250

    0.5
    0.4

    T

    -200 to 150
    150 to 400

    0.4
    0.5

    R

    -50 to 800
    800 to 1750

    0.8
    2.0

    S

    -50 to 850
    850 to 1750

    0.9
    2.0

    B

    100 to 1200
    1200 to 1800

    2.0
    3.0

    N

    -270 to 260
    260 to 1300

    0.5
    1.0

    E -50 to 1000 0.3
         

    Resolution

    0.1°C or 0.1°F

     

    Note: An additional correction representing the equivalent 1μV should be allowed for stray thermal emf effects (Temperature measure & simulation).

    PT100 Simulation
    14 set temperature points (°C) Accuracy
    -100, -50, -20, 0, 20, 50, 100, 200, 300, 400, 500, 600, 700, 800 0.1% of resistance value (typically 0.5 °C)
    PT100 Measure
    Range Accuracy Resolution
    200 to 700 °C, 2 wire. 0.2% of resistance value (typically 0.7 °C) 0.2°C (or 0.2°F)
    Millivolt Measure
    Range Resolution Accuracy Input Resistance

    0 to +/- 30mV

    10μV

    0.05% of full scale ±1 digit

    100kΩ

    Millivolt Source
    Range Resolution Accuracy Output Resistance
    0 to 8mV 0.5μV +/- 4μV 10Ω
    8 to 80mV 5μV 0.02% of full scale 10Ω
    Milliamp Measure
    Range Resolution Accuracy Input Resistance

    0 to 60mA

    20μA

    0.05% of full scale ±1 digit

    0.5Ω

    Milliamp Source
    Range Resolution Accuracy
    0 to 8mA 0.5μA +/-10μA

    8 to 80mA

    5μA

    0.02% of full scale

    Further Features and Functions
    Inching Three levels of increment, 0.1, 1 or 10 for °C/°F, and 1, 10, or 100 μV/ μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output.

    24V Process Loop Drive

    A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA
    (or a lower level if required)

    Memory recall & step

    10 memory locations for non-volatile storage of values. Manual & AutoStep, rate adjustable from 1 to 10 sec/step


    General Specifications

    Cold Junction Compensation Accuracy 0.2°C. Resolution 0.1°C.
    Operating Temperature -10 to 40°C (15 to 105°F)
    Connections Industry standard 4mm screw terminals

    Module Width

    295mm (primary or secondary console fitting)

    Optional Extras Calibration certificates traceable NPL and UKAS
  • Module

    7077

    Process Calibrator Module

    Options
    C177 Factory Calibration Certificate (NPL traceable)
    C139 UKAS Calibration Certificate (ISO 17025)