Pneumatic – The connection is made via two font panel
quick release connector ports. These are available for differential
pressure input (Hi and Lo ports).
Electrical - The 7066 is powered from an internal power
supply located in the console.
Optional: Vacuum regulator and rear panel connection.
The regulator is for vacuum and can be configured to supply
the differential pressure on the negative side of the 7066
differential pressure module. The regulator must in turn
be supplied by a primary high vacuum electric pump (not
supplied) – this should be connected via the console
rear. The level of vacuum achievable will depend on the
efficiency of the Primary Pump.
| Specifications |
| Range
|
Vacuum |
0.2 |
2 |
5 |
10 |
| Resolution (bar) |
.001 |
0.01mb |
.0001 |
0.001 |
0.001 |
| Acc % +/- 1 digit |
0.04 |
0.1 |
0.04 |
0.04 |
0.04
|
| Full Scale |
- |
199.99
mbar |
1.9999
bar |
5.000 |
10.000 |
| Temp Stability |
100ppm per degC |
| Units |
Bar, PSI, Kg/Cm2, inHg, mmHg,
KPa, cmWg. ** |
| Linearity |
0.02% of full scale |
| Max Static Pressure |
10bar |
| Sensor |
Piezo-resistive – stainless
steel diaphragm |
| Over press warning |
1.2 x range full scale –
audio and visual (on LCD) warning |
| Fittings |
Quick Release |
| |
** Specify on ordering: 5
pressure units max |
| General Information and
Options |
| Panel size |
Width 150 x Height 201mm |
| Order code |
7066 / Pressure range option
code |
| Pressure Options |
7100 = 200mbar range
7101 = 2bar range
7102 = 5bar range
7103 = 10bar range
7111 = Vacuum regulator |