Fully integrated with the 7051 Control Centre and EasyCal
the user can define pressure calibration procedures.
| Specifications |
| Range
|
Vacuum |
0.2 |
2 |
5 |
10 |
20 |
35 |
70 |
100 |
200 |
400 |
600 |
| Resolution (bar) |
.0001 |
0.01mb |
.0001 |
0.001 |
0.001 |
0.001 |
0.01 |
0.01 |
0.01 |
0.01 |
0.1 |
0.1 |
| Acc % +/- 1 digit |
0.04 |
0.1 |
0.04 |
0.04 |
0.04
|
0.04 |
0.04 |
0.04 |
0.04 |
0.1 |
0.1 |
0.1 |
| Temp Stability |
Less than 70 ppm per degC |
| Units |
Bar, PSI, kPa, Mpa, inWg,
cmWg, inHg, mmHg, Kg/cm2, atm** |
| Maximum Pressure |
2 x range. 1.5 x range for
20 bar and above |
| Sensor |
Piezo-resistive – stainless
steel diaphragm |
| Over press warning |
1.2 x range full scale –
audio and visual (on LCD) warning |
| Fittings |
Minimess (> 20 bar) and
Quick Release (< 20 bar) |
| |
** Specify on ordering: 5
pressure units max |
| General Information and
Options |
| Connections |
Quick release coupling for
0.2 to 20 bar ranges.
Minimess coupling for 35 to 600 bar ranges. |
| Panel size |
Width 96 x Height 201mm |
| Order code |
7065 / Pressure range option
code |
| Pressure Options |
7100 = 200mbar range.
|
7101 = 2bar range. |
7102 = 5bar range. |
| 7103 = 10bar range. |
7104 = 20bar range. |
7105 = 35bar range. |
| 7106 = 70bar range. |
7107 = 100bar range. |
7108 = 200bar range. |
| 7109 = 400bar range. |
7110 = 600bar range. |
7111 = Vacuum regulator. |
|