Features:

 

1mV – 220V Square Wave/DC

 

Frequency 0.1 – 100MHz

 

Time Markers 15s to 10ns

 

Fast Rise <300ps
  Front Panel or PC Virtual Control
  External reference input
  2 GHz Sweep Option
  Rubidium Option
  Current Probe Adaptor
RS232/USB/GPIB
Description:

The 5045 is a versatile, high accuracy calibrator capable of calibrating a wide range of oscilloscopes and timer/counters. It provides a wide range of outputs for amplitude, frequency, period and bandwidth. Amplitude calibration is achieved by a DC signal or 1kHz square-wave, ranging from 1mV to 220V (2V max for 50 ohm loads). Deviation up to ±9.99% allows fine adjustment of amplitude and direct read-out of error. Accurate frequencies are generated from a temperature controlled quartz crystal oscillator. Timing accuracy of 0.1ppm is suitable for most oscilloscopes and timer counters. A precise square-wave output provides a fast rise time of less than 300ps, which allows bandwidth testing up to 600MHz.

Simple Operation
Functions and ranges are easily accessed from the front panel. Increase and decrease keys per digit, are used to quickly set the output value. Deviation control then enables the user to finely adjust the output value as a percentage (+/-9.99%). All this information is shown on a clear, easy to read LED display.

Virtual Front Panel Software
Supplied as standard is Time Electronics’ windows based ‘Virtual Front Panel’ interface that allows the user to control the 5045 via a laptop or PC.

2 GHz Levelled sine-wave option
For precise bandwidth determination and frequency response analysis the 2 GHz option is available. The ability to sweep the frequency output from 100MHz to 2 GHz and adjust the amplitude to 0.5V, 1.0V and 1.5V pk-pk ensures accurate analysis of oscilloscope input amplifiers.

Current probe calibration
For calibration of oscilloscope current probes an external adaptor is available. This converts the 5045’s amplitude output to current and covers the range 0.1mA to 100mA pk-pk, 0.2% accuracy, DC or 1kHz.

Rubidium frequency reference
Enhanced timing performance is also available by specifying the rubidium frequency reference option (9762).
This option achieves timing accuracies required to calibrate high performance Timer/Counters to 1 part in 10¹º.

Calibration made easy
To automate the oscilloscope calibration process the 5045 can be controlled using Time Electronics’ EasyCal calibration software. This reduces calibration times, ensures consistent results and produces certificates to international quality standards.

 