High
specification oscilloscope/timer calibrator at an affordable
price
The 5041 is a versatile, high accuracy calibrator capable of
calibrating a wide range of oscilloscopes and timer/counters
to 2.2GHz. It provides a wide range of outputs for amplitude,
frequency, period and bandwidth.
Amplitude calibration is achieved by a DC signal or 1kHz square-wave,
ranging from 1mV to 200V (2V max for 50 ohm loads). Deviation
up to ±9.99% allows fine adjustment of amplitude and
direct read-out of error.
Accurate frequencies are generated from a temperature controlled
quartz crystal oscillator. Timing accuracy of 0.1ppm is suitable
for most oscilloscopes and timer counters. A precise square-wave
output provides a fast rise time of less than 300ps, which allows
bandwidth testing up to 600MHz.
Virtual front panel
The 5041 follows the modern trend of instrumentation that is
PC controlled. The Windows based ‘Virtual Front Panel’
is quick and easy to use.
Communication
with the unit is via RS232 (serial) or GPIB, which allows simple
integration with Time Electronics EasyCal Software. This also
provides the perfect solutionfor ATE systems.The 16-character
front panel display shows functions and
settings for user convenience.
2.2GHz
Levelled sine-wave option
For precise bandwidth determination and frequency response analysis
the 2.2GHz option is available. The ability to sweep the frequency
output from 10MHz to 2.2GHz and adjust the amplitude from 400mV
to 1.1V pk-pk ensures accurate analysis of oscilloscope input
amplifiers.
Current
probe calibration
For calibration of oscilloscope current probes an external adaptor
is available. This converts the 5041’s amplitude output
to current and covers the range 0.1mA to 100mA pk-pk, 0.2% accuracy,
DC or 1kHz.
Rubidium
frequency reference
Enhanced timing performance is also available by specifying
the rubidium frequency reference option (9762). This option
achieves timing accuracies required to calibrate high performance
Timer/Counters to 1 part in 10¹º.
Calibration made easy
To automate the oscilloscope calibration process the 5041 can
be controlled using Time Electronics’ EasyCal calibration
software. This reduces calibration times, ensures consistent