Power
Calibration |
|
DC Current
0.2 – 2.2A
2.2 - 22A |
Accuracy
0.03 + 500uA
0.05 + 6mA |
Compliance
5V
4V |
Resolution
100uA
1mA |
DC
Voltage
1- 22V
22 – 220V
220 – 1050V |
Accuracy
0.01 + 500uV
0.02 + 30mV
0.05 + 50mV |
Output
Current
20mA
20mA
10mA |
Resolution
100uV
1mV
10mV |
| Phase
45 to 99Hz
100Hz to 400Hz |
Accuracy
0.3 deg
1.0 deg |
Range
+/-90deg
+/-90deg |
Resolution
0.1 deg
0.1 deg |
| AC Current 45-400Hz
0.2 – 2.2A
2.2 - 22A |
Accuracy
0.1% + 2mA
0.1% + 20mA
|
Compliance
3.5V
3V
|
Resolution
100uA
1mA |
| AC
Voltage 45-400Hz
1 - 22V
22 – 220V
220 – 1050V |
Accuracy
0.03% + 2mV
0.06% + 30mV
0.08% + 90mV |
Output
Current
20mA
20mA
10mA |
Resolution
100uV
1mV
10mV |
| Power Factor
45 to 99Hz
100Hz to 400Hz |
|
Range
0.00 – 1.00
0.00 – 1.00 |
Resolution
0.001
0.001 |
The accuracy of the power is complex
and is determined by using a formula, which combines the
errors due to Voltage, Current, and Phase.
Power Acc (%) = SqrRt (Vacc^2 + Iacc^2 + Phase Correction^2)
Where Phase Correction (%) = 100x(1-Cos(Phase+PhaseAcc)/Cos
Phase).
The current and voltage terminals must be isolated. A current
transformer or clamp meter adaptor must be used if instrument
under test has a common negative. |