| The
5011 is primarily a programmable resistance
/ RTD source. With the ability to add DC Voltage, Thermocouple
Simulation and DC Current the 5011 is truly a versatile temperature
calibrator. High
accuracy resistance of 0.01% is offered as standard with 1mOhm
resolution from 20 Ohm to 1 kOhm.
Simple
front panel operation allows the user to quickly set the function
and output required. Using the jog / shuttle dial deviation
the user can finely adjust the output value as a percentage
(+/-9.99%). All this information is shown on a clear, easy to
read LED display.
As standard the 5011 uses a SCPI command structure
for programming. The 5011 also supports the older 9811/19/20
command set making it an ideal replacement for these models.
The DCV / Thermocouple option provides a DC
voltage source (+/- 20V) and simulates thermocouple types K,
J, T, R, S, N, E and B.
A DC Current option is also available which
sources up to 220mA. This makes the 5011 ideal for accurate
process control calibration.
Calibration
Made Easy
Connect the 5011 to a PC/Laptop (via RS233, GPIB or USB) installed
with Time Electronics EasyCal and automate the calibration process.
Increase speed of calibration and consistency of results, produce
calibration certificates and reports to ISO 9001 quality standards.
|